Posts

Showing posts from May, 2020

Usg Equipment Mindray DC-N3 Pro/DC-N3T/DC-N3S/DC-N3/DC-N3 Information and Troubleshooting Methods

en Mindray DC-N3. *Note: You can also use this artical for Mindray DC-N3 Pro/DC-N3T/DC-N3S. Basic Overview-            Its a basic Doppler machine. You can use this machine for basic application. You can also done 3D/4D. In this series DC-N3 is also available in vet model i.e use for veterinary use. It's a Linux base machine therefore you may have difficulty to install external devices such as printer, thermal printer etc i will share the list of supported printers and other supporting stuff. There are two unit in this machine as follows           1.Front-end Unit.           2.Back-end Unit.           3.Power Supply Unit. 1.Front-end Unit-           A)Probe Board-           ...

Usg Equipment Safety Precaution.

Basic Safety Precautions of Ultrasound Machine. *Note:Its basic Precautions for all machines not for dedicated machine. Basic Three types of Safety Precaution that are follows:           1)Electric Safety           2)Mechanical Safety           3)Personal Safety 1.Electric Safety-           a)Use of power plug of machine   and other external or peripherals devices such as printer, External monitor etc, that meets the power rating of it.           b)Always check and confirm the AC power of machine is disconnected while you cleaning, installing any hardware part and uninstall the machine.           c)Do not use the machine simultaneously with electrosurgical unit, high frequency therapy equi...

Usg Equipment Mindray DC-7 information and troubleshooting methods.

Mindray DC-7 Basic Overview-            It's a color  Doppler machine and it usable for whole body application like obs,  gyn, small parts, kidneys you can also use this machine for cardiac application.            Mindray machine basically coming with two types of Operating System one is Linux and  another is Windows in Windows OS Machine can use any type of printer but in Linux OS you only add the listed printer that list is available in the Linux base machine. Mindray DC-7 having Windows XP OS. .           its comes with LCD monitor with contrast and brightness controlling buttons. You can also adjust it with monitor support arm to position up and down. This system is comes with various input and output connector like USB, VGA output, DVI output,  Ethernet RJ45 connector, S-Video, BNC connector, Audio Stereo output etc. It has user-f...

Philips hd11xe photo gallery.

Image
  (A)  Signal Processing Board This board is used for signal processing of ultrasound its located in usg section without this board machine will not boot. This board transfer signal to PC section its only one board that communicate with PC side via token ring/i pass cable. it also process the data that collected from TR Boards. (B) TR(Transmission and Received) Board This Board is used for transmit and received signal from probe and send it to signal processing board. In machine we found multiple tr boards as per machine model and its configuration. In some machine transmit and received board both are different and separate function done of transmit and received. (C) Graphics Card This card is play important role in machine it can install externally also it available in some machine internally or onboard. We cannot see the high quality images/screen without this. it available various model as per machine model or software version. (D) Bac...

How to find out Philips hd11xe restart error solution..???

Image
Title -We discuss about the restart error solution for Philips hd11xe ultrasound Equipment. Ans - we all are know the hd 11xe machine of Philips and it's common issue of this machine. We have a some ways to find out the solution of this problem. First we divide the machine in two sections 1)PC section 2)Usg section 1)PC section-     If restart error occurred and you suspect the pc section is faulty then you need to check and boot machine only with pc section (motherboard,hdd, monitor) if restart error occurred with only pc section then you need to replace or reload software. If machine will boot up normally and shows the probe sector screen then probably problem is different or not in pc section.(for checking pc section you need to graphic card output is mandatory otherwise it shows new graphic card found massage). 2)Usg section-     In usg section we have          a) Power Supply    ...